Standard |
Item |
IEC 60601-1-2 Test Levels |
Remarks |
|
IEC 60601-1-2 |
Electrostatic Discharge |
Contact: ± 8kV Air: ± 2kV, ± 4kV, ± 8kV, and ± 15kV |
No performance degradation observed. |
|
|
Radiated Immunity |
80MHz to 2.7GHz, 10V/m; @ 80%, 1kHz and Table 9 of Clause 8.10 |
No performance degradation observed. |
|
|
Electro Fast Transients/Bursts |
AC Mains: 2kV, 100kHz PRF DC Mains: NA |
No performance degradation observed. |
|
|
Surge |
AC Mains: ± 0,5 kV and ± 1 kV Line to Line, ± 0,5 kV, ± 1 kV and ± 2 kV Line to Ground DC Mains: NA IO Lines: NA |
No performance degradation observed. |
|
|
Conducted Immunity |
AC Mains: 150kHz to 80MHz, 3 Vrms outside ISM and amateur radio bands, 6 Vrms in ISM and amateur radio bands: @ 80%, 1kHz IO Lines: 150kHz to 80MHz, 3 Vrms outside ISM and amateur radio bands, |
No performance degradation observed. |
|
|
|
6 Vrms in ISM and amateur radio bands: @ 80%, 1kHz DC Mains: NA |
|
|
|
Magnetic Immunity |
30A/m, 50Hz and 60Hz |
No performance degradation observed. |
|
|
Voltage Dips and Interrupts |
0% for 0.5 cycle (at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°), 0% for 1 cycle (at 0°),70% for 25 cycles (at 0°), and 0% for 250 cycles. (Tested at 230VAC and 100VAC) |
No performance degradation observed. |
Comments
0 comments
Please sign in to leave a comment.